top of page

What causes cavitation or air in the pump, and how do I purge it correctly?

January 16, 2026

System type: Liquid Chromatography (LC)

System type: Liquid Chromatography (LC)

Pump, system pressure

Cavitation and Entrained Air in HPLC/UHPLC Pumps: How to Diagnose the Cause and Purge Correctly

Overview

Cavitation and gas bubbles in the pump inlet path are among the most frequent drivers of unstable pressure, noisy baselines, retention time drift, and gradient delivery errors in HPLC/UHPLC and LC–MS workflows. Although the symptoms often look similar on the pressure trace, the underlying mechanisms differ:

  • Cavitation occurs when the local inlet pressure drops below the solvent’s vapor pressure, forming vapor pockets that collapse as pressure recovers inside the pump.

  • Entrained air refers to actual gas bubbles pulled into the pump through leaks, poor degassing, empty reservoirs, or agitation/outgassing.

Both conditions reduce effective displacement per stroke, disrupt check-valve sealing, and can create severe flow pulsation.

Core operating principle:
Always establish a fully flooded, bubble-free inlet path and purge to waste through a low-resistance route before reconnecting any high-backpressure component (column, inline filters, MS interface restrictions).

What You Will See When Air or Cavitation Is the Problem

Typical Symptoms

  • Pressure oscillations, “sawtooth” patterns, or intermittent step drops/spikes

  • Audible clicking, chatter, or knocking from check valves

  • Inconsistent flow, variable retention times, and poor gradient reproducibility

  • Visible bubble trains in transparent inlet tubing (or pump head viewing windows on some instruments)

  • Detector baseline spikes/noise; in LC–MS, unstable spray and TIC fluctuations

  • Prime/degasser alarms, slow priming, or failed prime routines

Why It Matters

Air in the pump head behaves like a compressible spring. The pump may “build pressure” without delivering correct flow, leading to:

  • Erratic pressure readings and unstable baseline

  • Wrong gradient composition (especially in low-pressure mixing systems)

  • Increased mechanical wear on seals and check valves due to repeated valve chatter

Root Causes: Where Air and Cavitation Come From

1) Solvent Supply and Degassing Problems

  • Solvent reservoir low or empty; inlet frit partially exposed

  • Degasser turned off, overloaded, leaking, or failing

  • Warm solvents or warm lab conditions increasing vapor pressure

  • Volatile solvents (notably acetonitrile) outgassing after agitation or temperature change

  • Excessive sparging or vigorous mixing creating persistent bubbles rather than removing dissolved gas

2) Plumbing and Hardware Faults on the Inlet Side

  • Loose inlet fittings that admit air under suction (often no visible liquid leak)

  • Cracked or partially clogged inlet frits/filters (salt crystals, particulates, microbial films)

  • Long, narrow inlet tubing that increases suction losses and reduces net positive suction head

  • Contaminated or worn check valves; worn piston seals allowing internal leakage and instability

  • Purge valve or purge restrictor partially blocked (prevents effective purging)

  • Low-pressure proportioning valve leaks or sticking (quaternary systems), which can pull air or deliver inconsistent composition

3) Operating Condition Triggers

  • Priming at very high flow with viscous or cold solvents (inlet starvation)

  • Altitude (lower atmospheric pressure) combined with warm solvents (higher vapor pressure)

  • Abrupt solvent changes that reduce wetting (starting “water-first” on a dry pump head)

  • Shutdown with buffers followed by restart without proper flushing (salt crystallization in valves/seals)

Quick Diagnostic Checklist (Fast Localization)

Run these checks before disassembly:

  1. Solvent bottles
    Are inlet frits fully submerged?
    Are filters visibly clean and not clogged?

  2. Degassing
    Is the degasser enabled and assigned correctly to channels?
    Do you see persistent bubbles downstream of the degasser?

  3. Inlet fittings
    Tighten inlet-side fittings (hand-tight plus appropriate wrenching per SOP).
    Watch for microbubbles appearing at ferrules or junctions.

  4. Purge function
    Confirm purge valve opens and produces a continuous stream to waste.
    If purge flow is weak or intermittent at typical priming flow, suspect purge restriction or inlet starvation.

  5. Channel isolation
    Prime channels one at a time into waste. A single “problem channel” often points to an inlet leak, clogged frit, or proportioning valve issue.

  6. Mechanical cues
    If the pressure fluctuation frequency matches pump stroke frequency and you hear chatter, check valves and trapped air in the head become primary suspects.

The Correct Way to Purge an HPLC/UHPLC Pump (General Procedure)

Preparation: Set Up for Success

  • Choose a wetting solvent first: methanol, isopropanol, or a 50:50 IPA:water mixture is often effective because it wets seals and displaces bubbles more readily than high-water phases.

  • Create a low-resistance path to waste:
    Open the purge/prime valve.
    Bypass the column and any restrictive inline hardware.

  • Set safe pressure limits and ramp flow gradually, not abruptly.

Step-by-Step Purge Sequence

  1. Confirm solvent supply integrity
    Fresh solvent, correct bottle venting, inlet frit submerged.

  2. Open purge valve fully
    Route purge outlet to a waste container you can visually monitor.

  3. Prime at an appropriate flow rate
    Analytical-scale systems commonly use higher prime flows than method flows, but use values consistent with your instrument’s guidance and inlet capacity.
    The key objective is a steady, bubble-free stream, not a maximum flow number.

  4. Prime each channel individually
    Purge each line long enough to eliminate bubbles and stabilize the stream.
    Channel-by-channel priming prevents one aerated line from contaminating the entire mixing path.

  5. If bubbles persist: flood the pump head
    Stop flow.
    Use syringe-priming via the designated prime port (if available) to push several mL of solvent into the head and displace trapped gas.
    Resume purging.

  6. Address check valve chatter
    Gently tap the check valve housing (light, controlled tapping) to help release a trapped bubble.
    If instability remains, remove and clean/replace check valves per your SOP. Persistent chatter after purging often indicates contamination or wear.

  7. Stabilize under modest backpressure
    After purge is bubble-free, close the purge valve.
    Apply modest controlled backpressure (or reconnect a low-restriction test setup) and confirm the pressure trace is smooth and repeatable.

  8. Reintroduce the column gradually
    Only reconnect the column and bring the system to method conditions once pressure is stable and bubble-free.

What “Good” Looks Like After Purging

  • Stable pressure trace without abrupt dips/spikes

  • No audible knocking

  • Bubble-free inlet lines

  • Normal prime behavior and restored gradient reproducibility

Purging by Pump Architecture

Quaternary Pumps (Low-Pressure Mixing)

  • Prime each channel independently with the purge valve open.

  • A single channel with persistent bubbles often indicates:
    inlet leak on that line
    clogged inlet frit
    proportioning valve timing/leak issue

  • Keep suction-side tubing short and appropriately sized to reduce inlet pressure drop.

Binary Pumps (High-Pressure Mixing)

  • Use built-in prime routines where available (often optimized for dual-head evacuation).

  • Trapped air around outlet check valves can cause severe pulsation—confirm check valve function after purging.

UHPLC and Microflow Systems

  • Use lower prime flows to avoid inlet starvation.

  • Use an IPA or methanol wetting step first; microbubbles are harder to dislodge at very low internal volumes.

  • Allow extra time for stabilization because small bubbles can persist longer in narrow bores.

Preparative Pumps

  • Verify inlet tubing is large-bore and inlet filtration is not restrictive.

  • Prime at moderate flow first, then step up only after stable flow to waste is established.

Degassing Practices That Actually Prevent Recurrence

Vacuum Degassing

  • Confirm correct channel routing and avoid exceeding the degasser’s practical throughput per channel.

  • Persistent bubbles after the degasser can indicate membrane aging, vacuum weakness, or internal leaks.

Helium Sparging

  • Use gentle sparging to strip dissolved gases without foaming.

  • Avoid aggressive bubbling that can create stable microbubbles and worsen pump performance.

  • Do not sparge during analysis in ways that disturb composition or temperature.

Reservoir Handling

  • Minimize headspace when practical and use appropriate venting.

  • Avoid major temperature swings between solvent preparation, storage, and instrument use.

Preventing Air and Cavitation Long-Term

Solvent Hygiene

  • Replace inlet frits/filters regularly.

  • Filter mobile phases and avoid microbial growth in aqueous lines.

  • Equilibrate solvents to room temperature before use.

Optimize Inlet Plumbing

  • Keep inlet lines short and adequately wide-bore.

  • Use robust inlet tubing and ensure fittings are seated correctly.

  • Eliminate unnecessary connectors on the suction side.

Operational Habits

  • Prime after solvent changes, bottle swaps, or long idle periods.

  • Wet the system with methanol/IPA before switching to high aqueous mobile phases if the system is dry.

  • Use seal wash when running buffers; flush salts before shutdown.

Proactive Service

  • Clean/replace check valves and piston seals on a maintenance cycle.

  • Ensure purge valves and purge lines are not partially blocked.

LC–MS Considerations (Why This Becomes More Visible)

Air and pulsation upstream can translate directly into:

  • spray instability and signal fluctuation

  • TIC spikes and poor quantitative repeatability

  • increased sensitivity to post-column outgassing (pressure drops can release dissolved gases)

Maintaining bubble-free flow and stable pump operation is therefore not optional in LC–MS—it directly affects ionization stability and data quality.

Why Cavitation Happens (Practical Physics)

Cavitation occurs when absolute inlet pressure falls below solvent vapor pressure. Anything that lowers inlet pressure or raises vapor pressure increases risk:

  • long/narrow tubing, clogged frits, high suction demand (high pump speed)

  • warm solvents, warm lab conditions

  • lower ambient pressure (altitude)

Mitigation is straightforward: reduce suction losses, ensure degassing, keep solvents at stable temperature, and prime at conditions the inlet can support.

Summary

Cavitation and entrained air produce pressure noise, unstable flow, gradient errors, and detector/MS instability by disrupting pump displacement and check valve function. Reliable correction requires a structured purge method: establish a flooded inlet path, open a low-resistance route to waste, wet the system with an appropriate solvent, prime channels individually, flood the head if needed, and verify stability under modest backpressure before reconnecting the column. Prevent recurrence by controlling degassing, maintaining inlet plumbing, and servicing check valves/seals proactively.

Relevant Articles

ChemITrust AI & Lab Services

bottom of page