What causes negative peaks or baseline dips in HPLC?
January 19, 2026

System type: Liquid Chromatography (LC)
Instrument Part: Baseline, System signal, blank runs
Negative Peaks and Baseline Dips in HPLC: Causes, Diagnostics, and Corrective Actions
Executive Overview
Negative peaks (downward deflections) and baseline dips in HPLC occur when the detector signal temporarily decreases relative to the established baseline. This is not “random behavior”—it is almost always the result of a detector-response contrast between the injected plug (sample/diluent) and the surrounding mobile phase, or a transient system disturbance (mixing, temperature, bubbles, autozero timing).
Rule of thumb: A negative peak most often means the injected plug or eluting component produces a lower detector response than the mobile phase under the current conditions.
This guide explains why negative peaks happen, how the mechanism differs by detector type (UV/DAD, RI, ELSD/CAD, fluorescence, electrochemical), and provides a practical diagnostic workflow and fix list to eliminate negative dips during isocratic and gradient HPLC.
What Does a “Negative Peak” Mean in HPLC?
A negative peak indicates one (or more) of the following is true at that moment in the run:
The mobile phase absorbs more (UV) than the plug passing through the detector.
The refractive index (RI) of the plug is lower than the mobile phase.
Detector processing (e.g., reference wavelength subtraction) is overcorrecting baseline behavior.
A transient event (bubbles, mixing ripple, temperature change) temporarily reduces the detector signal.
Mechanisms by Detector Type
1) UV/Vis and DAD/PDA: Most Common Causes of Negative Peaks
Negative peaks in HPLC-UV/DAD are frequently explained by mobile-phase absorbance being higher than the sample plug or diluent, especially at low wavelengths.
Mobile Phase Absorbance Greater Than the Injected Plug
At low wavelengths, mobile-phase constituents can absorb strongly (examples in your text include TFA at 214 nm and phosphate near 200–210 nm).
When a plug with lower absorbance enters the flow cell, the signal dips negative.
Wavelength Too Close to Solvent Cutoff
Operating near solvent cutoff increases baseline absorbance and sensitivity to small disturbances:
MeOH ~205 nm
ACN ~190 nm
Water ~190 nm
Near these regions, minor composition or temperature changes can generate negative excursions.
DAD/PDA Reference-Wavelength Subtraction Artifacts
An inappropriate reference band (e.g., 360–450 nm) can overcorrect baseline changes (especially during gradients), producing negative peaks.
Mis-matched bandwidths or selecting a reference region that changes during the run can invert small signals.
Gradient Composition Changes
Baseline shifts as %B changes.
If the background decreases faster than the sample absorbance increases, you can observe local negative dips.
Sample-Induced pH/Composition Perturbation
Injecting strong acid/base/buffer can locally change the mobile-phase matrix absorbance and cause negative deflections.
Strong Diluent Effects
Example: injecting pure water into a mobile phase with higher absorbance (e.g., aqueous with additive or mixed solvents) can create a negative front.
2) Refractive Index (RI): Contrast and Temperature Sensitivity
Negative peaks in RI are expected whenever the sample/diluent RI is lower than the mobile phase.
Refractive Index Contrast
If the plug has a lower RI than the mobile phase, the detector responds negatively.
Gradient elution inherently perturbs RI, often producing pronounced dips.
Temperature Mismatch
Small differences (your text: ±0.2–0.5 °C) between sample, column, and detector can create RI transients.
Flow/Pressure Fluctuations
Pump pulsation and degassing failures produce RI dips or spikes.
3) CAD / ELSD: Solvent Volatility and Baseline Re-Equilibration
Composition Transients
Rapid solvent changes (gradient steps) can generate short negative deflections as the nebulizer/dryer baseline re-equilibrates.
Gas Flow / Temperature Instability
Momentary cooling or gas flow changes can lower baseline signal.
4) Fluorescence: Quenching and Inner-Filter Effects
Quenching
A co-eluting matrix or diluent can quench baseline fluorescence (from background impurities), producing negative dips.
Inner-Filter Effects
Strong UV-absorbing plugs reduce excitation light reaching the cell, lowering emission transiently.
5) Electrochemical: Polarity and Baseline Offset
If an analyte undergoes reduction at a potential set for oxidation (or vice versa), it can appear as a negative deflection.
Offset drift or reference electrode instability can intensify the effect.
Method-Related Root Causes
Negative peaks and baseline dips become more likely when any of these conditions apply:
Wavelength selection too low (<210–220 nm) while using UV-absorbing additives (e.g., TFA, phosphate).
Sample diluent mismatch versus initial mobile phase (%B, pH, ionic strength).
Additives not matched in both channels (A vs B) or insufficient equilibration in gradients.
Injection volume too large relative to column volume, causing plug-driven perturbation.
Inadequate degassing and dissolved gas effects (especially UV and RI).
DAD reference subtraction misconfigured.
Adsorption/desorption of ion-pair reagents/surfactants causing system peaks.
High organic starting conditions with aqueous diluent plug leading to focusing/defocusing artifacts.
Instrument and Hardware Contributors
Detector Autozero Timing
Autozero just before/at injection can lock the baseline during a transient.
When conditions stabilize, the return can appear negative.
UV Lamp Energy / Stray Light / Flow Cell Condition
Aging lamp and unstable optics increase baseline instability.
Dirty flow cells increase baseline variability; reference subtraction can invert small signals.
Temperature Control Gaps
Unthermostatted column or detector increases baseline fluctuations and negative dips.
Contamination and Desorption
Adsorbed UV-absorbing residues can desorb during gradients, producing system peaks that may be negative or positive.
Pump Mixing and Proportioning Ripple
Short-term composition excursions can appear as dips, especially in gradients.
Microbubbles
UV/DAD: bubble reduces effective path length or scatters light → negative spike.
RI: bubbles produce strong excursions.
Diagnostic Workflow: Step-by-Step Isolation
1) Run a Blank Gradient (No Injection)
If dips appear without injection, suspect:
Gradient baseline behavior
Mixing/proportioning ripple
Additives and solvent effects
Thermal instability
2) Perform Diluent-Only Injections (n = 3)
If negative dips occur with diluent:
Diluent mismatch, injection volume, or RI contrast is likely.
3) Isocratic Hold Test at Initial %B
Hold isocratic at starting conditions and inject diluent.
If negative peaks disappear: gradient-related artifacts are likely.
4) Wavelength Sweep
Repeat at 254 nm or >230 nm.
If dips vanish: low-wavelength solvent/additive absorbance is implicated.
5) DAD Reference Off / Adjust
Disable reference subtraction or set a stable reference configuration (example from your text: Ref 380–450 nm, BW 100 nm).
If dips disappear: reference subtraction artifacts are causal.
6) Temperature Control Test
Thermostat column and detector (example: 30–40 °C).
Reduction in dips supports thermal mismatch or RI sensitivity.
7) Injection Volume Study
Reduce injection volume stepwise.
Smaller dips with smaller volume indicate plug perturbation effects.
8) Degassing and Bubble Checks
Verify vacuum degassing (or helium sparging if used).
Tap detector cell to evaluate microbubbles.
Confirm detector backpressure device is present (example: ~5–10 bar on UV cell).
9) Column Bypass Test
Connect detector directly to mixer.
If dips persist: pump/mixer/degassing are likely.
If dips vanish: column/diluent interaction or additive adsorption/desorption is likely.
10) Organic Solvent Swap (Diagnostic Only)
Switch ACN ↔ MeOH (with proper revalidation planning).
Useful for probing solvent cutoff contributions.
Corrective Actions That Eliminate Negative Peaks
Optimize UV/DAD Detection Settings
Increase wavelength to ≥230–254 nm when using UV-absorbing additives.
Avoid problematic reference subtraction; disable reference or ensure reference band is stable during gradients.
Avoid operating too close to solvent cutoff.
Match Sample Diluent to Mobile Phase
Prepare sample in a diluent that matches initial %B, pH, and ionic strength.
Reduce injection volume if plug effects are evident.
Use on-column focusing appropriately by controlling starting %B.
Stabilize Gradient and Additives
Ensure UV-absorbing additives are present at identical concentration in both A and B.
Premix when proportioning accuracy is limited.
Equilibrate sufficiently before injections (example guidance in your text: ≥10–20 column volumes).
Improve Degassing and Flow Stability
Verify degasser performance; replace membranes if necessary.
Use fresh, filtered, degassed solvents.
Maintain inlet frits and purge thoroughly.
Use a backpressure device to reduce outgassing effects.
Control Temperature
Thermostat column and detector.
Allow sample vials to equilibrate to ambient or stabilize tray conditions.
Clean the Flow Path and Detector Hardware
Flush flow cell using a solvent sequence compatible with your system (example from your text: IPA → water → mobile phase).
Replace contaminated tubing/tees; clean mixer if needed.
Manage Adsorptive Additives and System Peaks
Reduce or avoid ion-pair reagents where possible.
Consider volatile acids (example: formic acid 0.1%) instead of TFA at low wavelengths if method constraints allow.
For required ion-pairing: increase equilibration and run blank injections before samples.
RI / ELSD / CAD Adjustments
RI: avoid gradients where possible; if unavoidable, manage acquisition timing and stability.
ELSD/CAD: stabilize gas flow and temperatures; avoid abrupt gradient steps; use shallow ramps.
Understanding System Peaks During Gradients
System peaks arise from column interactions with mobile-phase components (salts, acids, ion-pair agents) that are retained and later released.
Key identifiers
Appear in blank runs
Shift with gradient conditions
Can be negative if the released plug decreases detector response
Mitigation
Match additives in A and B
Extend equilibration
Reduce strongly adsorbing components
Consider guard columns to stabilize surface chemistry
Quantitation and Data Handling Guidance
Do not “solve” negative peaks by integration tricks unless the root cause is understood.
Exclude early solvent/system peaks using time events where appropriate.
Validate baseline stability under controlled variations (wavelength, temperature, injection volume).
If negative peaks are meaningful for a specific study (e.g., fluorescence quenching), document and standardize conditions.
Quick Troubleshooting Checklist
Is UV wavelength ≥230–254 nm when using absorbing additives?
Is sample diluent matched to initial %B, pH, ionic strength?
Are degassing, lamp energy, and flow cell cleanliness verified?
Is DAD reference subtraction configured correctly (or off)?
Was the system equilibrated for ≥10–20 column volumes?
Are column and detector temperature controlled?
Do blank and diluent-only injections show the same dips?
Summary
Negative peaks and baseline dips in HPLC most commonly occur when a transient plug—diluent, sample matrix, or released mobile-phase component—has a lower detector response than the surrounding mobile phase. The dominant drivers are low-wavelength UV operation with absorbing additives, diluent mismatch, gradient baseline behavior, DAD reference subtraction artifacts, temperature/RI sensitivity, microbubbles, and mixing instability. A structured workflow using blank gradients, diluent-only injections, wavelength and reference tests, injection volume reduction, degassing checks, and column bypass quickly isolates the root cause and directs targeted corrective action.