top of page

Why did my LC-MS signal suddenly drop or disappear?

January 20, 2026

System type: Liquid Chromatography Mass Spectrometry (LC-MS)

System type: Liquid Chromatography Mass Spectrometry (LC-MS)

Instrument Part: Signal and Detector

LC–MS Signal Suddenly Dropped or Disappeared: Root Causes, Diagnostics, and Fixes (ESI/APCI, Triple Quad, QTOF, Orbitrap)

Keywords: LC–MS no signal, LC-MS signal loss, LC-MS sensitivity drop, ESI spray unstable, TIC flatline, MRM not detected, ion source contamination, divert valve to waste, ion suppression, vacuum fault, nitrogen gas issue

A sudden LC–MS signal drop (or complete disappearance) is almost always traceable to one of five areas: LC flow/plumbing, ion source/interface, MS tuning/acquisition, vacuum/detector hardware, or sample/matrix ion suppression. The fastest path to resolution is to isolate LC vs MS using a reference standard and a short, structured set of tests.

Scope and Symptom Definition

This troubleshooting guide addresses sudden loss of LC–MS signal, including:

  • No peaks, drastically reduced intensity, or a flat baseline (TIC or extracted channels)

  • Unstable spray, intermittent signal, or frequent source alarms

  • Applicable to ESI and APCI sources across triple quadrupole, QTOF/Orbitrap, and single quadrupole instruments

  • Focused on real instrument/method causes (not data processing/integration artifacts)

Start Here: Rapid Triage to Locate the Failure

1) Determine Whether the Problem Is LC-Side or MS-Side

Perform a direct infusion of a known reference standard (tuning mix or a stable analyte standard):

  • Infusion normal → MS is fundamentally working; suspect LC flow, divert valve, plumbing, chromatography, or sample matrix

  • Infusion low/absent → suspect ion source, gas supply, vacuum, detector, interlocks, or wrong tune/method state

2) Check Instrument Status Messages and Interlocks Immediately

Review the instrument status panel/logs for:

  • HV disabled, source door open, missing cover, spray interlock

  • Vacuum faults, pump errors, or high background pressure

  • Gas supply low, nitrogen generator fault, dew point out of spec

These conditions can cause the instrument to auto-protect by disabling high voltage or limiting acquisition performance.

High-Probability Root Causes (and How to Prove Them)

1) Ion Source and Interface Failures (ESI/APCI)

Common Causes

  • Spray instability or loss (ESI): inadequate nebulizer/drying gas, incorrect temperatures, emitter/capillary blockage, contaminated cone/orifice, incorrect voltages

  • APCI sensitivity collapse: corona discharge instability, gas/temperature mismatch, contamination on the probe

  • Source HV off due to interlock or method state

  • Wrong polarity or wrong ionization mode (ESI vs APCI) in the method

  • Divert valve routing to waste during the analyte elution window

Diagnostics That Work Fast

  • Confirm source HV = ON and the correct polarity is selected

  • Verify nitrogen pressure, flow, and temperature setpoints

  • Inspect emitter tip / spray needle / transfer capillary for salt crust or particulates

  • Inspect cone/orifice for contamination films or deposits

  • Confirm divert valve timing: temporarily route the whole run to MS to test

Corrective Actions

  • Clean cone/orifice and relevant interface parts per vendor SOP

  • Replace/clean emitter and transfer capillary if restricted

  • Restore known-good source parameters (gas flows, temps, voltages)

  • Correct polarity/mode and verify the tune state used for acquisition

  • Fix divert valve program so the analyte window goes to MS, not waste

2) LC Flow, Degassing, and Plumbing Issues (Silent Signal Killers)

Common Causes

  • No flow or partial flow to the source (leak, empty bottle, pump cavitation)

  • Degasser failure introducing microbubbles → unstable spray and TIC spikes

  • Blocked frit/column/inline filter → unstable flow or altered split behavior

  • Post-column restriction or a change in tubing ID/length starving the MS

  • Unintentional split ratio change (or partial blockage in split capillary)

Diagnostics

  • Check pump pressure trace:
    Sudden pressure drop → leak/no flow
    Pressure oscillations/ripple → bubbles/cavitation/check valves
    Sudden pressure increase → blockage/restriction

  • Confirm actual flow at the outlet (where feasible) and inspect for leaks at unions

  • Prime lines until flow is stable and bubble-free

  • Trace the entire path from column outlet → divert → source inlet (do not assume it matches yesterday)

Corrective Actions

  • Re-prime and purge solvent lines; confirm degasser status and vacuum behavior

  • Fix leaks; remake fittings; remove salt crystallization on connections

  • Replace inline filters/guard column; address blockages

  • Restore the correct post-column tubing and split configuration

  • If needed for stability, run briefly at lower flow and ramp up while watching pressure and TIC

3) Mobile Phase Composition and “Instant Ion Suppression”

A method can appear unchanged while the LC–MS response collapses due to a solvent or additive change.

Common Suppression Triggers

  • Non-volatile buffers (phosphate, sulfate) or high ionic strength

  • Detergents/surfactants (SDS, Tween)

  • High TFA (or sudden change in acid/base strength)

  • New solvent lot containing impurities or higher background contaminants

Diagnostics

  • Inject a clean standard prepared in a clean, volatile-compatible diluent

  • Compare response using a freshly prepared mobile phase with volatile additives

  • If a standard suddenly works after switching mobile phase lots, the root cause is likely composition/contamination, not the MS hardware

Corrective Actions

  • Replace non-volatile components with volatile buffers (ammonium formate/acetate)

  • Minimize or remove TFA when possible

  • Standardize solvent preparation, labels, and lot tracking

  • Flush the system to remove old additive films before resuming sequences

4) Chromatography Changes That Hide the Peak (Even When the MS Works)

High-Frequency Failure Modes

  • Retention shifts move the analyte into a divert-to-waste window

  • Scheduled MRM windows are too tight after a column/temperature change

  • Column plugging or collapse causes broad, low peaks that look like “no signal”

  • Injection solvent mismatch causes breakthrough, splitting, or nonreproducible elution

Diagnostics

  • Temporarily send the entire run to MS (disable divert timing) and check for signal return

  • Widen scheduled retention windows (or temporarily disable scheduling)

  • Compare a standard in a clean matrix vs your sample

  • Check for sudden backpressure increase (blockage) or peak shape degradation

Corrective Actions

  • Update divert valve timing to match real elution

  • Widen scheduled windows and re-lock them after stabilization

  • Replace guard/column if pressure and peak shape indicate plugging

  • Match injection solvent strength to initial mobile phase; reduce injection volume if needed

5) Acquisition Method or Tune Configuration Errors

Common Causes

  • Wrong MRM transitions, wrong mass range, or disabled transitions

  • Incorrect collision energy or polarity mismatch

  • Too many MRMs → long cycle time → reduced dwell → apparent intensity collapse

  • Wrong tune file loaded (lens voltages or interface settings not appropriate)

  • Detector gain/EM settings limited or disabled

Diagnostics

  • Confirm precursor/product m/z, CE, polarity, and retention scheduling

  • Confirm scan range and resolution mode match your method intent

  • Load a known-good tune and compare critical voltages and lens settings

  • Check cycle time and dwell times against peak widths

Corrective Actions

  • Restore known-good method and tune

  • Reduce event count or improve scheduling to bring cycle time under control

  • Confirm detector settings are enabled and within expected ranges

6) Vacuum, Detector, and Hardware Faults

Common Causes

  • Vacuum instability after venting, pump faults, or leaks

  • Source door/cover interlocks disabling HV

  • Electron multiplier (EM) aging or auto-limiting behavior

Diagnostics

  • Check rough and high-vacuum readings and pump status

  • Confirm covers and interlocks are fully seated

  • Review detector gain/EM voltage and noise trends

Corrective Actions

  • If vacuum is out of spec or unstable, pause operation and follow your service escalation path

  • Clean source and interface, replace O-rings/gaskets where appropriate

  • Follow vendor qualification steps for EM performance and replacement when indicated

Step-by-Step Diagnostic Workflow (Designed for Speed)

Step 1 — Confirm MS Health (Direct Infusion)

Infuse a known standard at a steady low flow.

  • Normal signal → proceed to LC-side checks

  • Weak/no signal → focus on source/gas/vacuum/tune/detector

Step 2 — Verify Source, Gas, and HV

  • Nitrogen supply pressure and dryness/purity are within spec

  • HV is on; correct polarity/mode selected

  • Clean/inspect cone/orifice and emitter/capillary

Step 3 — Confirm LC Flow and Degassing

  • Prime/purge until bubble-free

  • Inspect for leaks and restrictions

  • Review pressure trace for cavitation or blockages

Step 4 — Validate Divert Valve and Plumbing

  • Confirm analyte window goes to MS

  • Temporarily route everything to MS to rule out valve timing errors

Step 5 — Verify Acquisition Method and Tune

  • Confirm transitions, scan ranges, scheduling, cycle time

  • Reload known-good tune and compare settings

Step 6 — Evaluate Matrix Suppression vs True Instrument Loss

  • Inject a clean standard vs the real sample

  • If standards work but samples fail, suppression/cleanup is the main target

Step 7 — Maintenance/Service Escalation

  • Clean source/optics/ion optics per SOP

  • If vacuum/detector faults persist, escalate to service

Best Practices to Prevent Sudden LC–MS Signal Loss

  • Run a daily system suitability check with a reference compound

  • Track mobile phase composition, pH, lots, and preparation notes

  • Avoid non-volatile buffers; favor ammonium acetate/formate

  • Keep scheduled MRM windows with safety margins and re-verify after any column/gradient change

  • Standardize injection solvent strength and volume

  • Maintain a documented cleaning cadence for source and interface components

Summary

A sudden LC–MS signal drop is most commonly caused by ion source/interface contamination, gas/HV/interlock conditions, LC flow or degassing failures, divert valve misrouting, method/tune misconfiguration, matrix ion suppression, or vacuum/detector faults. The fastest isolation strategy is a direct infusion check, followed by targeted verification of source conditions, LC flow stability, divert valve routing, and acquisition settings.

Relevant Articles

ChemITrust AI & Lab Services

bottom of page