Why does my baseline shift after injection even when peaks still appear?
January 19, 2026

System type: Liquid Chromatography (LC)
Instrument Part: Baseline, System signal, blank runs
Baseline Shifts After Injection in HPLC While Peaks Remain: Causes, Diagnostics, and Fixes
Executive Overview
A baseline shift immediately after injection—while analyte peaks still elute normally—usually indicates a system or detector response to the injected plug (diluent, temperature, matrix, refractive index, or composition), rather than a problem with the analytes themselves. This symptom is common in LC-UV/DAD, RI, fluorescence, ELSD/CAD, and LC–MS, and it can occur under both isocratic and gradient conditions.
The key to solving it is to classify the baseline behavior and then isolate whether the shift follows the injection plug, the detector, the pump/mixing/degassing, or the column and thermal environment.
What the Symptom Means
When the baseline shifts after injection but peaks remain, you are typically seeing one of these baseline patterns:
Step change (offset): abrupt upward or downward shift shortly after injection or solvent front
Drift: gradual baseline movement over minutes after the injection event
Transient ripple or spike: damped oscillation or brief excursion following the injection pulse
Because the injected plug reaches the detector after the system’s delay volume, the timing of the shift (immediate vs delayed) is a powerful clue.
Rapid Triage Questions
Use these questions to narrow the likely cause within minutes:
Is the method isocratic or gradient?
What detector is in use (UV/DAD, RI, FL, ELSD/CAD, MS)?
Does a blank injection of diluent reproduce the baseline shift?
Is the shift positive or negative? Is it immediate or delayed by the system delay volume?
Does the shift disappear when the column is bypassed?
Does the shift scale with injection volume?
Are autosampler and column temperature controlled and stabilized?
Do you see simultaneous pressure changes or pump ripple during injection?
Most Common Root Causes in Liquid Chromatography
1) Sample and Diluent Mismatch
Diluent mismatch is one of the most frequent drivers of post-injection baseline offsets.
Typical mechanisms
Diluent stronger than initial mobile phase (higher organic, different pH/ionic strength) causes a discontinuity in absorbance and refractive index and can flush retained background material from the column head.
Diluent contains chromophores at the detection wavelength (for example, UV-active components in some solvents) and shifts the UV baseline.
Diluent temperature differs from the column/mobile phase temperature, creating density/RI differences (very strong in RI, but also observable in UV/DAD and MS through spray stability).
Common clue
The baseline shift is tightly tied to the solvent front timing and often scales with injection volume.
2) Mobile Phase and Gradient Optical Effects (UV/DAD)
For UV/DAD, baseline steps and offsets often trace to absorbance differences between solvent components and additives.
Typical mechanisms
Solvent A and B differ in absorbance at the detection wavelength; a plug with slightly different composition can create a step even in “isocratic” conditions if the plug deviates from bulk composition.
UV-absorbing additives (e.g., strong low-UV absorbers) or impurities present in one reservoir can create steps when local concentration changes.
DAD features such as reference wavelength, baseline correction, or overly wide bandwidth can amplify composition-driven offsets.
Common clue
The magnitude changes significantly with wavelength or with changes to reference/bandwidth settings.
3) Detector-Specific Artifacts (By Detector Type)
UV/DAD
Lamp not fully warmed up can drift, and the injection perturbation can make the drift appear as a step.
Contaminated flow cell or a trapped bubble can shift baseline after the mechanical shock of injection.
RI
RI responds strongly to any composition or temperature mismatch; gradients inherently shift the RI baseline and are typically incompatible without special compensation hardware.
Even small temperature differences between sample, column, and detector can cause prominent positive/negative deflections.
Fluorescence
Matrix quenching or enhancement from the injected plug changes baseline offset.
Surfactants/solvents can alter fluorescence background behavior.
ELSD/CAD
Nonvolatile components in diluent (salts, PEG, detergents) increase scattering/background after injection.
Drift tube temperature or nebulizer gas instability amplifies baseline shifts.
MS (ESI/APCI)
Injection plugs with different ionic strength, surface tension, or organic fraction can shift TIC baseline by changing spray current and desolvation efficiency.
Strong plugs can suppress or enhance signal transiently, even when analyte peaks remain visible later.
4) Pump, Mixing, and Degassing Issues
Injection is a system disturbance, and it can expose marginal stability in the fluidics.
Typical mechanisms
Inadequate degassing leads to microbubbles that dislodge at injection and create baseline steps or drift.
Proportioning valve wear, check valve sticking, or incorrect compressibility settings can create composition ripple that is triggered or worsened by injection pulses.
In low-delay-volume systems, the plug reaches the detector quickly, making slight mixing errors more visible.
Common clue
Baseline shift coincides with visible pressure ripple or changes in baseline behavior after priming/purging.
5) Column, Flow Path, and Thermal Effects
A column that is not fully equilibrated or a system with thermal instability can show baseline offsets specifically after injection.
Typical mechanisms
Incomplete equilibration after gradients leaves the stationary phase transient; the next injection displaces local conditions and shifts baseline.
Temperature transients across the column or detector cause viscosity/RI changes at injection time.
Fouled guard column or column bleed can release a background “slug” upon injection.
Common clue
Increasing equilibration time reduces the shift, or bypassing the column removes the symptom.
6) Autosampler and Needle Wash Artifacts
Autosampler solvents and injection mechanics can introduce composition discontinuities.
Typical mechanisms
Needle wash solvent incompatible with initial mobile phase or containing UV-absorbing residues shifts baseline.
Overfilled partial-loop injections create discontinuities.
Air bubbles in the loop or poor metering create detector perturbations.
Large injection volume relative to column volume exaggerates the shift.
Common clue
The shift correlates with wash events, injection mode, or decreases when injection volume is reduced.
Gas Chromatography Considerations (If Applicable)
Baseline offsets after injection can also occur in GC:
Solvent plug overload or mismatch can create a baseline offset near the solvent front.
Septum bleed, liner contamination, or column bleed can present as a post-injection “shift.”
Inlet pressure pulsing or split changes at solvent vent close can cause FID excursions; for MS, transient source pressure changes can alter baseline.
Fast, Conclusive Diagnostic Experiments
1) Blank Injection Series
Inject diluent only. If the shift repeats, the diluent/injection plug is implicated.
Inject initial mobile phase as the “sample.” If the shift disappears, composition mismatch is the primary cause.
2) Injection Volume Scaling Test
Reduce injection volume stepwise.
If the baseline shift scales linearly with volume, it is strongly plug-driven.
3) Column Bypass Test
Bypass the column with a union and repeat a blank injection.
Shift persists → detector, pump/mixing, degassing, or diluent
Shift disappears → column interaction, equilibration, guard/column contamination, or thermal effects
4) Standardize or Swap the Diluent
Prepare sample in initial mobile phase and re-test baseline behavior.
5) Temperature Stabilization Test
Match autosampler tray and column oven temperatures and observe whether shift magnitude decreases.
6) Detector-Specific Checks
UV/DAD: disable reference wavelength; adjust bandwidth; compare 210 vs 254 nm to test solvent/additive absorbance effects.
RI: confirm strong thermal stability and verify no gradient use.
ELSD/CAD: verify drift tube temperature and nebulizer gas stability; compare neat diluent vs salt-containing diluent.
MS: monitor spray current; use early divert to waste and compare TIC baseline with/without divert.
7) Pump/Degassing Tests
Purge/prime all channels, verify degasser vacuum, and observe pressure ripple during injection.
Persistent ripple supports check valve or mixing instability.
8) Equilibration Extension
Increase re-equilibration by 5–10 column volumes and confirm baseline is stable before injecting.
9) Flow Cell Integrity Check (UV)
Inspect/clean UV flow cell; flush with isopropanol/water; ensure no trapped air.
Corrective Actions That Solve Baseline Shifts After Injection
Composition and Injection Control
Match sample diluent to initial mobile phase (organic fraction, buffer type, ionic strength, pH).
Reduce injection volume or use full-loop injections with appropriate loop sizing to minimize composition slugs.
Add an initial isocratic hold to buffer the injection plug effects when appropriate.
Use a weaker, non-absorbing needle wash; separate strong wash from the injection stream where possible.
For LC–MS, divert early eluent to waste through the solvent front.
Detector-Specific Fixes
UV/DAD
Select wavelengths where the mobile phases are transparent and avoid overly low UV when additives dominate absorbance.
Turn off or tune reference wavelength; narrow bandwidth where appropriate; ensure lamp warm-up.
Clean/replace flow cell and eliminate bubbles via proper degassing and stable backpressure.
RI
Avoid gradients; strictly match diluent to mobile phase and stabilize detector/column temperature.
Minimize injection volume and prevent temperature deltas between sample and system.
ELSD/CAD
Remove nonvolatile components from diluent.
Stabilize drift tube temperature and nebulizer gas.
Avoid abrupt gradient steps and maintain adequate equilibration.
MS (ESI/APCI)
Minimize nonvolatile salts and matrix load.
Maintain stable source/gas conditions; use early divert where necessary.
Consider approaches that stabilize spray conditions (e.g., managing early eluent and solvent front effects).
System and Method Robustness Improvements
Verify degassing performance, prime lines, and service check valves and seals as needed.
Stabilize autosampler and column temperatures and avoid large temperature deltas.
Increase re-equilibration after gradients and confirm stable backpressure before injection.
Refresh guard columns and replace fouled columns releasing background material.
Add baseline stability checks to system suitability where appropriate.
Impact on Quantitation and Integration
Baseline offsets can bias integration, especially if global baselines are applied.
Use local baseline windows and peak-specific baselines.
Avoid forcing a zero baseline when a verified offset exists.
Apply a controlled autozero only when it does not clip analyte signal and is performed under stable conditions.
Symptom Pattern Examples
These examples map a baseline pattern to a likely mechanism and corrective action:
Positive UV step immediately post-injection that scales with injection volume
Likely stronger organic (or otherwise mismatched) diluent; fix by matching diluent and reducing injection volume.Negative UV step at low wavelength during a gradient
Likely solvent absorbance mismatch; select a higher wavelength or tune DAD reference and verify solvent quality.Large symmetric deflection on RI baseline in isocratic mode
Likely RI or temperature mismatch; match diluent and stabilize temperatures.TIC baseline sag during the first minute of ESI-MS
Likely matrix-related suppression; use early divert and reduce nonvolatile salts.
Quick Checklist
Does a diluent-only injection reproduce the baseline shift?
Does injecting initial mobile phase remove the shift?
Does the shift scale with injection volume?
Does bypassing the column eliminate the shift?
Are autosampler and column temperatures stable and matched?
For UV/DAD: are wavelength, bandwidth, and reference settings appropriate?
For MS: is early divert to waste implemented when needed?
Is degassing verified and pressure ripple controlled?
Summary
A baseline shift after injection while peaks remain is most often caused by the injected plug and its impact on absorbance, refractive index, temperature, or matrix effects—rather than a problem with analyte elution. Rapid diagnostics (blank injections, injection volume scaling, column bypass, wavelength/reference tests, and temperature/degassing control) isolate the root cause quickly. Corrective actions center on matching diluent to initial mobile phase, optimizing injection and detector settings, stabilizing temperatures, ensuring robust degassing and pump stability, and confirming sufficient column equilibration.